Maintaining the Purged Packed Inlet 5
Maintaining Your GC 71
10 Firmly press the new septum or Merlin Microseal into the 
fitting. 
11 Install the septum retainer nut or Merlin cap and 
finger-tighten. Tighten the septum retainer nut until the 
C-ring is about 1  mm above the nut. 
12 Restore the analytical method.
13 Reset the septum counter. 
CAUTION
Overtightening the septum nut can cause contamination.
1 mm