86 Maintaining Your GC
5 Maintaining the Purged Packed Inlet
5
Tighten the inlet column nut 1/4  turn with a wrench.
6 After conditioning, remove the column from the inlet. 
7 Simultaneously insert the column into the inlet and detector 
fittings but do not force it. It may be necessary to start the long 
end of the column in the inlet at an angle to clear the oven floor.
8 Withdraw the column 1 to 2  mm from both the inlet and 
detector. Finger-tighten both column nuts.
9 Tighten both column nuts 1/4  turn with a wrench.
10 Configure the new packed column (make sure either column 
length or diameter is zero).
11 Establish a flow of carrier gas and purge as recommended by 
the packing manufacturer. Generally:
• 20 to 30  mL/min for 2-mm id glass or 1/8-inch od metal 
columns
• 50 to 60  mL/min for 4-mm id glass or 1/4-inch od metal 
columns
12 Restore the analytical method.
• For NPD, immediately set the bead voltage to 0.0.
13 After the GC becomes ready, wait 10 minutes then ignite the 
detector flame or adjust offset on the NPD bead.
WARNING
Do not use hydrogen as the carrier for conditioning! It could vent 
into the oven and present an explosion hazard.
CAUTION
Overtightening the column nut or forcing it to bottom in either the 
inlet or detector may shatter the column.