1c-4
Operation messages
Front Panel Operation Messages
0
2
1
Copying registers from 8647/8
This message is displayed while the memory registers are being copied from the
instrument to the 83301A Memory Interface.
0
2
2
Press
SAV to copy memory TO 8647/8
This message is displayed when the⇑ key is pressed on an 83301A Memory Interface
connected to the instrument. Press the
SAV key to copy memory registers that had been
saved in the memory interface into the instrument.
0
2
3
Copying registers to 8647/8
This message is displayed while the memory registers are being copied from the 83301A
memory Interface to the instrument.
0
2
4
Invalid data in Memory Interface
This message is displayed when the instrument detects that the 83301A Memory
Interface does not contain valid memory register data. Try copying a memory register
into the memory interface first, then initiate a copy from the memory interface to the
instrument again.
0
2
5
Communication failure: copy aborted
This message is displayed when the instrument is not able to successfully copy memory
registers between the instrument and the 83301A Memory Interface. This message will
be displayed if the cable connecting the instrument to the primary interface is
disconnected during the copy process.
0
2
6
Memory copy was successful
This message is displayed when the instrument has successfully copied the memory
registers between the instrument and the 83301A Memory Interface. When this
message is displayed, the copy process is complete and you can disconnect the memory
interface from the instrument.