EasyManua.ls Logo

Agilent Technologies 8648A - Required Test Equipment

Agilent Technologies 8648A
481 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8-3
Performance Tests
Required Test Equipment
Required Test Equipment
• 8903B Option 051 Audio Analyzer
— Distortion accuracy: ±1 dB
— Residual distortion: −80 dB at 80 kHz BW
— 30 kHz low-pass filter
— AC level accuracy: ±4%
— CCITT weighting filter
• 8902A Measuring Receiver
— FM accuracy: ±2% of reading ±1 digit
— AM accuracy: ±2% of reading ±1 digit
— Range: 150 kHz to 1300 MHz
— Filter: 300 Hz high-pass
— Filter: 15 kHz low-pass
— Detectors Peak+
• 8563E RF Spectrum Analyzer
— Frequency range 1 MHz to 4000 MHz
— Relative level accuracy: ±2 dB
(harmonic and spurious measurements)
— 100 Hz digital resolution BW filter (required for power accuracy)
— Displayed average noise:
100 kHz: −110 dBm
1 MHz to 10 MHz: −130 dBm
10 MHz to 4 BHz: −134 dBm
— RF input VSWR: 1:1.5
— External timebase input
• 8663A Synthesized Signal Generator
— Frequency range: 11.5 MHz to 2500 MHz
— Output level: ±4 Hz at 4000 MHz
— Meets 8663A single-sideband phase noise specifications

Table of Contents