7-20
Adjustments
Manual Adjustments
Output level: Frequency Extension Calibration (8648B/C/D Only)
Description
This adjustment creates the slope and offset calibration data for the output calibration.
The adjustment will not let you store away any calibration data unless all of the
calibration data points are run.
Required Test Equipment
Power Meter
Power Sensor
Procedure
Figure 7-19. Frequency Extension Calibration Test Setup
1. Connect the equipment as shown above.
2. Preset all of the equipment.
3. Follow the instructions as they are displayed on the PC.