EasyManuals Logo

Agilent Technologies 8648A User Manual

Agilent Technologies 8648A
481 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #261 background imageLoading...
Page #261 background image
7-5
Adjustments
Manual Adjustments
Internal Reference Oscillator Adjustment
Use this procedure to adjust the internal timebase reference DACs. The internal reference
oscillator is adjusted with two DACs, one for coarse tuning and one for fine tuning. Using
the two DACs, the internal reference oscillator can be adjusted to the resolution of the
frequency counter used.
NOTE This manual adjustment can be replaced by performing the automated Time
Base DAC calibration instead.
Required Test Equipment
5316B Frequency Counter
5071A primary Frequency Standard
Procedure
1. Turn off power to the signal generator.
2. Set the rear panel TIMEBASE ADJUST switch to on (1). (See Figure 7-2.)
Figure 7-2. Timebase Adjust Switch Location
3. Turn on power to the signal generator and wait for the self-tests to run.
4. Connect the equipment as shown in Figure 7-3.(The DUT is the signal generator under
test.)
Figure 7-3. Internal Reference Oscillator Adjustment Setup

Table of Contents

Other manuals for Agilent Technologies 8648A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8648A and is the answer not in the manual?

Agilent Technologies 8648A Specifications

General IconGeneral
BrandAgilent Technologies
Model8648A
CategoryTest Equipment
LanguageEnglish

Related product manuals