7-6
Adjustments
Manual Adjustments
5. Follow the instructions on the signal generator’s display and adjust the knobs until the
frequency counter reads 100 MHz within 1 Hz resolution.
6. When the adjustment is complete, turn off the signal generator.
7. Set the rear-panel TIMEBASE ADJUST switch to off (0).