9-2
Supplemental Verification Tests
Required Test Equipment List
Required Test Equipment List
• 3458A Digital Multimeter (DMM)
Range: −50 V to +50 V
Functions: DC and AC
DC accuracy: 0.01%
Input impedance: > 10MΩ
AC range: 1 kHz
AC accuracy: ±0.1%
• 5350B Frequency Counter
Frequency range: 10 Hz to 20 GHz
Frequency accuracy: ±4 Hz at 4000 MHz
(Includes reference accuracy and counter accuracy)
• 8493A Option 010 10 dB Attenuator
No substitutes