1b-39
Operation Reference
Function
Memory
The memory keys allow you to save instrument settings into memory registers and recall
the registers in a numeric sequence.
Up to 10 register sequences can be defined (0 through 9). A sequence can contain up to 100
registers (00 through 99). There are a total of 300 registers available in the instrument.
The registers can be used in the sequences in any combination (such as 10 sequences of 30
registers each, or 3 sequences of 100 registers each) as long as the total does not exceed
300 registers. It is not possible to have all 10 sequences each contain 100 registers as that
would be 1000 registers. (If Option 1EP is present, there are a total of 70 registers
available.)
Memory key entries are automatically terminated after you enter the last digit. Register
key entries (
SAV, DEL, and REG), require two digits. Sequence key entries (SEQ) require
one digit.