8-4
Performance Tests
Required Test Equipment
• 5350B Frequency Counter
— Frequency range: 10 Hz to 20 GHz
— Frequency accuracy: ±4 Hz at 4000 MHz
(Includes reference accuracy and counter accuracy)
• MDC-174 RF Mizer (M/A Com Inc.)
— Frequency range (RF and LO ports): 1 MHz to 2800 MHz
— Frequency range (IF port): 1 MHz to 2000 MHz
— Maximum input (LO port): +26 dBm
• MDC-164 RF Mizer (M/A Com Inc.) 8648C/D only
— Frequency range (RF and LO ports): 500 MHz to 9000 MHz
— Frequency range (IF port): 10 MHz to 2000 MHz
— Maximum input (LO port): +24 dBm
• 438 Power Meter
— Instrumentation accuracy: ±0.5%
— Power reference accuracy: 0.95
• 8481D Option H70 Power Sensor
— Frequency range: 100 kHz to 4.2 GHz
— Power range: −70 dBm to −20 dBm
—Maximum SWR:
— 100 kHz to 300 kHz: 1:1.2
— 300 kHz to 2 GHz: 1:1.15
— 2 GHz to 4.2 GHz: 1:1.4
— Power linearity (−30 to −20 dBm): ±1%
— Cal factor accuracy: ≤ 1.6%
• 8482A Power Sensor
— Frequency range: 100 kHz to 4.2 GHz
— Power range: −30 dBm to +13 dBm
—Maximum SWR:
— 100 kHz to 300 kHz: 1:1.6
— 300 kHz to 1 MHz: 1:1.12
— 1 MHz to 2 GHz: 1:1.1
— 2 GHz to 4.2 GHz: 1:1.1