5c-4  
Service Error Messages
515   Freq Ext level cal restore failure
Description   After detecting a failure in comparing calibration data between RAM and 
the calibration ROM, a data restore was attempted unsuccessfully.
Cause   There is either a failure in writing data to RAM or in reading data from 
the calibration ROM.
What To Do   The calibration ROM can be verified by replacing the A10 (freq ext) 
assembly. The calibration ROM is contained in the assembly.
If the failure is still present after replacing the A10 assembly, replace the 
A3 assembly, which contains the RAM.
516   Config cal restore failure
Description   After detecting a failure in comparing calibration data between RAM and 
the calibration ROM, a data restore was attempted unsuccessfully.
Cause   There is either a failure in writing data to RAM or in reading data from 
the calibration ROM.
What To Do   Replace the A3 assembly. The A3 assembly has a continuity failure. The 
calibration ROM is on the A3 assembly with the RAM.
517   Output prelevel cal restore failure
Description   After detecting a failure in comparing calibration data between RAM and 
the calibration ROM, a data restore was attempted unsuccessfully.
Cause   There is either a failure in writing data to RAM or in reading data from 
the calibration ROM.
What To Do   The calibration ROM can be verified by replacing the A6 assembly. The 
calibration ROM is contained in the assembly.
if the failure is still present after replacing the A6 assembly, replace the 
A3 assembly, which contains the RAM.
518   Atten comp loss cal restore failure
Description   After detecting a failure in comparing calibration data between RAM and 
the calibration ROM, a data restore was attempted unsuccessfully.
Cause   There is either a failure in writing data to RAM or in reading data from 
the calibration ROM.
What To Do   Replace the A3 assembly. The A3 assembly has a continuity failure. The 
calibration ROM is on the A3 assembly with the RAM.