8-138
Performance Tests
8648D Test Record
NOTE If the automated Power Level Accuracy performance test is performed, the
RF Level Accuracy performance test is not required.
Table 8-87. RF Level Accuracy Performance Test - Part 1
Frequency
(MHz)
Amplitude
(dBm)
Limits (dBm)
Uncertainty
(±dB)Lower Measured Upper
31312 140.22
10 9.0 11.0 0.22
4 3.0 5.0 0.22
5.9 6.9 4.9 0.22
15.9 16.9 14.9 0.23
25.9 26.9 24.9 0.23
35.9 36.9 34.9 0.23
45.9 46.9 44.9 0.23
55.9 56.9 54.9 0.24
65.9 66.9 64.9 0.24
75.9 76.9 74.9 0.25
85.9 86.9 84.9 0.25
95.9 96.9 94.9 0.25
96.0 97.0 95.0 0.25
105.9 106.9 104.9 0.26
106.0 107.0 105.0 0.26
115.9 116.9 −114.9 0.27
116.0 117.0 115.0 0.27
125.9 126.9 124.9 0.28
126.0 127.0 125.0 0.29
127.0 128.0 126.0 0.29