Loading...
9-10
Supplemental Verification Tests
9 kHz RF Level Accuracy Supplemental Verification Test
Table 9-10. 8648C 9 kHz RF Level Accuracy Supplemental Verification Test
Power
Level
(dBm)
Limits (kHz)
Uncertainty
(±dB)Lower Measured Upper
10 596 841 0.21
0 188 265 0.21
10 59 84 .021
20 19 27 0.21
30 6.0 8.0 0.21
40 1.9 2.7 0.21
50 0.566 0.894 0.25
60 0.179 0.283 0.25
70 0.055 0.094 0.39
Table 9-11. 8648D 9 kHz RF Level Accuracy Supplemental Verification Test
Power
Level
(dBm)
Limits (kHz)
Uncertainty
(±dB)Lower Measured Upper
10 596 841 0.21
0 188 265 0.21
10 59 84 .021
20 19 27 0.21
30 6.0 8.0 0.21
40 1.9 2.7 0.21
50 0.566 0.894 0.25
60 0.179 0.283 0.25
70 0.055 0.094 0.39

Table of Contents