9-10
Supplemental Verification Tests
9 kHz RF Level Accuracy Supplemental Verification Test
Table 9-10. 8648C 9 kHz RF Level Accuracy Supplemental Verification Test
Power
Level
(dBm)
Limits (kHz)
Uncertainty
(±dB)Lower Measured Upper
10 596 841 0.21
0 188 265 0.21
−10 59 84 .021
−20 19 27 0.21
−30 6.0 8.0 0.21
−40 1.9 2.7 0.21
−50 0.566 0.894 0.25
−60 0.179 0.283 0.25
−70 0.055 0.094 0.39
Table 9-11. 8648D 9 kHz RF Level Accuracy Supplemental Verification Test
Power
Level
(dBm)
Limits (kHz)
Uncertainty
(±dB)Lower Measured Upper
10 596 841 0.21
0 188 265 0.21
−10 59 84 .021
−20 19 27 0.21
−30 6.0 8.0 0.21
−40 1.9 2.7 0.21
−50 0.566 0.894 0.25
−60 0.179 0.283 0.25
−70 0.055 0.094 0.39