5 Maintenance
Cleaning a clogged capillary
Agilent Triple Quadrupole LC/MS System User Guide 118
Cleaning a clogged capillary
This EMF item relies on the rough vacuum reading to determine a clog in the
capillary/capillary. If the rough vacuum reading falls below a value of 1.4 Torr, the
EMF item will be flagged.
Schedule
When the rough vacuum reading flags an error.
Equipment List
Clean powder-free nitrile gloves
LC/MS-grade isopropanol
Capillary (new, if replacing)
Steps
1 Put on clean, powder-free nitrile gloves.
2 Remove and replace the capillary with a new unit, or reinsert the capillary after
cleaning and drying.
3 Monitor the rough vacuum level to ensure that it does not fall below 1.4 Torr.