5 Maintenance
Cleaning the spray chamber - weekly
93 Agilent Triple Quadrupole LC/MS System User Guide
5 Attach the calibrant bottle onto the fixed bottle cap. Turn the calibrant bottle
counterclockwise to tighten.
CAUTION
The bottle only needs to be snug. Do not over-tighten the bottle. A leaky bottle
cap can prevent the bottle from pressurizing and can lead to the evaporation of
the contents.
CAUTION
The tuning mixes are not interchangeable. Failure to change the tuning mix
when the source or inlet assembly is changed can result in miscalibration of
the instrument and errors in mass assignments.
Cleaning the spray chamber - weekly
Perform this maintenance weekly or whenever symptoms indicate that
contamination exists in the spray chamber and normal daily cleaning does not
correct the problem.
Tools needed:
Clean powder-free nitrile gloves
Clean lint-free cloth
LC/MS-grade isopropanol
LC/MS-grade water
Clean wash bottle
4000 grit abrasive paper
To perform weekly cleaning of the spray chamber,
1 Put on clean powder-free nitrile gloves.
2 Disconnect the nebulizer sample line and sample nebulizer gas tubing from
the nebulizer.
3 Disconnect all cables and tubings from the instrument.
4 Open the latch of spray chamber.
5 Open the spray chamber.
6 Put the source under a fume hood.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Triple Quadrupole LC/MS and is the answer not in the manual?

Agilent Technologies Triple Quadrupole LC/MS Specifications

General IconGeneral
BrandAgilent Technologies
ModelTriple Quadrupole LC/MS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals