5 Maintenance
Checking and filling the calibrant bottle - weekly
Agilent Triple Quadrupole LC/MS System User Guide 92
Checking and filling the calibrant bottle - weekly
Schedule
Perform this procedure monthly or weekly if you tune the instrument frequently,
or if the calibrant has expired.
Equipment List
APCI calibrant
APCI-L tuning mix
ESI-L tuning mix
MMI-L tuning mix
Reference mix
Clean powder-free nitrile gloves
Steps for checking calibrant levels
1 Check to make sure that enough tuning mix is present to immerse the end of
the intake tube.
2 If the tuning mix level is within a few millimeters of the end of the intake tube,
refill the calibrant bottle.
Steps for refilling the calibrant bottle
1 Rinse the calibrant bottle once with water, then twice with ACN before refilling.
CAUTION
Dispose of rinsing agents appropriately.
2 Label the calibrant bottle with the tuning mix expiration date.
3 Refill the calibrant bottle with the appropriate tuning mix.
4 Put the intake tube into the calibrant bottle as you lift the calibrant bottle into
position.
CAUTION
Do not touch the intake tube unless you are wearing clean gloves.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Triple Quadrupole LC/MS and is the answer not in the manual?

Agilent Technologies Triple Quadrupole LC/MS Specifications

General IconGeneral
BrandAgilent Technologies
ModelTriple Quadrupole LC/MS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals