3Tuning
Calibrating the LC/TQ (Checktune)
Agilent Triple Quadrupole LC/MS System User Guide 44
Calibrating the LC/TQ (Checktune)
6475A and Ultivo
A checktune can be run with the following ion sources: ESI, AJS ESI, MMI, and
APCI.
1 In the Method Editor window, click the TQ tab.
2 Click Tune > Autotune in the left pane.
3 Click Request tune control.
This button locks control of the TQ instrument. When tune has control of the
TQ instrument, a single sample run or a worklist cannot be started and the
method editor is locked.
4 Select a polarity.
If Both is selected, the system will automatically run in the positive mode
followed by the negative mode.
5 Click Checktune instrument.
When clicked, tune setpoints are applied. If an action is not selected within
the timeout period, the system will go back to the method setpoints.
To stop the tune during operation, click
6 When the tune completes, review the report(s).
7 Click Release tune control to release control of the TQ instrument.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Triple Quadrupole LC/MS and is the answer not in the manual?

Agilent Technologies Triple Quadrupole LC/MS Specifications

General IconGeneral
BrandAgilent Technologies
ModelTriple Quadrupole LC/MS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals