5 Maintenance
Source Maintenance
Agilent Triple Quadrupole LC/MS System User Guide 126
Source Maintenance
Changing to a source that uses a different tuning mix
Schedule
Every time you change a source that uses a different tuning mix.
Equipment List
None
Steps
1 Right-click the TQ device pane and select Calibrant.
2 Run the calibrant for 5 minutes.
3 Perform a checktune. See “Calibrating the LC/TQ (Checktune)” on page 44.
4 If the checktune results are not acceptable, run autotune on a supported
tuning source. See “Calibrating the LC/TQ (Autotune)” on page 48.
5 When the instrument is ready, load or create the method to use with the
source.
6 Change the method settings for the new source.
Installing and removing sources
Schedule
Every time you change a source that uses a different tuning mix.
Before you begin:
Do a complete autotune (see “Calibrating the LC/TQ (Autotune)” on
page 48). If an autotune is not available, do a checktune (see “Calibrating the
LC/TQ (Checktune) on page 44).
If the tune report does not show good results, tune the instrument. If needed,
change the source to one that supports autotune.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Triple Quadrupole LC/MS and is the answer not in the manual?

Agilent Technologies Triple Quadrupole LC/MS Specifications

General IconGeneral
BrandAgilent Technologies
ModelTriple Quadrupole LC/MS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals