5 Maintenance
Routine Tasks
91 Agilent Triple Quadrupole LC/MS System User Guide
Routine Tasks
Cleaning the spray chamber - daily
Perform this maintenance daily or at the end of each shift or anytime you suspect
carryover contamination from one sample or analysis to another.
NOTE
Use the weekly cleaning procedure if symptoms of contamination persist or if the
spray shield or capillary cap show significant discoloration that cannot be
removed by the regular, daily cleaning.
Tools needed:
Clean, powder-free nitrile gloves
Clean lint-free cloth
LC/MS-grade isopropanol
LC/MS-grade water
Clean wash bottle
To perform daily cleaning of the spray chamber,
1 Put on clean powder-free nitrile gloves.
2 Lower the drying gas temperature to the minimum level.
3 For APCI/multimode, lower the vaporizer heater temperature to the minimum
level.
WARNING
Make sure the source is cooled before you clean the spray shield.
4 In a clean wash bottle, prepare a 50:50 mix of LC/MS-grade isopropanol and
LC/MS-grade water.
5 Open the spray chamber.
6 Dampen a clean lint-free cloth with a mixture of LC/MS-grade isopropanol and
LC/MS-grade water.
7 Wipe the spray shield and the area around the spray shield.
8 Close the spray chamber.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Triple Quadrupole LC/MS and is the answer not in the manual?

Agilent Technologies Triple Quadrupole LC/MS Specifications

General IconGeneral
BrandAgilent Technologies
ModelTriple Quadrupole LC/MS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals