5 Maintenance
Replacing the electron multiplier horn (Ultivo only)
121 Agilent Triple Quadrupole LC/MS System User Guide
Replacing the electron multiplier horn (Ultivo only)
Schedule
When sensitivity is poor and autotune consistently sets the detector gain to its
maximum value.
Equipment List
Clean, powder-free nitrile gloves
Long-nose pliers
Electron multiplier horn
CAUTION
Take care when you perform this procedure. Damages that you cause during
maintenance are not covered under warranty. If you need assistance, contact
Agilent support.
Steps
1 Remove the left side cover.
2 Remove the detector horn carrier.
3 Disconnect the signal wire from the electron multiplier horn.
4 Open the retaining clip.
5 Remove the electron multiplier horn.
6 Install the new electron multiplier horn.
7 Close the retaining clip.
8 Connect the signal wire to the pin on the electron multiplier horn.
9 Install the detector horn carrier.
10 Install the left side cover.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Triple Quadrupole LC/MS and is the answer not in the manual?

Agilent Technologies Triple Quadrupole LC/MS Specifications

General IconGeneral
BrandAgilent Technologies
ModelTriple Quadrupole LC/MS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals