EasyManuals Logo

Agilent Technologies Triple Quadrupole LC/MS User Manual

Agilent Technologies Triple Quadrupole LC/MS
135 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #128 background imageLoading...
Page #128 background image
5 Maintenance
Agilent Jet Stream (AJS ESI) source
Agilent Triple Quadrupole LC/MS System User Guide 130
Agilent Jet Stream (AJS ESI) source
Removing
1 Disconnect the nebulizer sample line and sample nebulizer gas tubing.
2 Disconnect the heating cable.
3 Disconnect the sheath gas line at the connector and unplug it.
4 Disconnect the AJS HV cable delivering AJS nozzle voltage.
5 Disconnect the Multimode cable delivering AJS nozzle voltage.
6 Open the latch of the spray chamber.
7 Open the latch of spray chamber.
WARNING
The spray chamber operates at very high temperature. Do not continue until
the spray chamber is cool.
8 Open the spray chamber.
9 Remove the ion source by lifting it along the axis of its hinges.
WARNING
The spray shield can be hot. Be careful not to burn yourself when you remove
the spray shield.
10 If you are changing to a different source type, remove the spray shield
Installing
1 Install the spray chamber.
2 Close the spray chamber.
3 Connect the MultiMode cable delivering AJS nozzle voltage.
4 Connect the heater cable.
5 Connect the sample nebulizer gas tubing.
6 Connect the nebulizer sample line.
7 Connect the reference nebulizer sample line and sample nebulizer gas tubing.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Triple Quadrupole LC/MS and is the answer not in the manual?

Agilent Technologies Triple Quadrupole LC/MS Specifications

General IconGeneral
BrandAgilent Technologies
ModelTriple Quadrupole LC/MS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals