3Tuning
Scheduling a Checktune
53 Agilent Triple Quadrupole LC/MS System User Guide
3 Select Checktune in the left pane.
The right pane shows the information for scheduling a checktune.
NOTE
The polarity is chosen for this instrument using the functionality “Configure Tune
(6495D only)” on page 46.
4 Click the Scheduling slider to switch Scheduling on.
5 Select either Weekly or Monthly.
6 Select other options to indicate how often to schedule the checktune.
7 Click Save.