4 Basic Operation
Turning off the instrument
Agilent Triple Quadrupole LC/MS System User Guide 62
3 In MassHunter Data Acquisition, right-click the TQ icon in the Instrument
Status window and select Vent to vent the system. Click Yes to continue.
When venting the system, the following occurs:
The spray chamber high voltages, the drying gas heater, the quad heaters,
the collision cell gas, the nebulization flow, the detector, and other lens
voltages are turned off.
The drying gas flow is set to Standby.
If the APCI source is installed, the vaporizer heater is also turned off.
If the installed source is an Agilent Jet Stream source, the sheath gas
heater is also turned off.
The rough pump turns off when the turbo speed is below a certain
threshold.
4 Wait until the instrument completely vents.
5 Close the MassHunter Data Acquisition program and shut down the
computer.
6 Press the power switch located in the lower left front corner of the instrument.
7 Unplug the main power cable behind the service panel.
This action prevents the turbo pump components from overheating.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Triple Quadrupole LC/MS and is the answer not in the manual?

Agilent Technologies Triple Quadrupole LC/MS Specifications

General IconGeneral
BrandAgilent Technologies
ModelTriple Quadrupole LC/MS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals