4 Basic Operation
Turning off the instrument
Agilent Triple Quadrupole LC/MS System User Guide 62
3 In MassHunter Data Acquisition, right-click the TQ icon in the Instrument
Status window and select Vent to vent the system. Click Yes to continue.
When venting the system, the following occurs:
The spray chamber high voltages, the drying gas heater, the quad heaters,
the collision cell gas, the nebulization flow, the detector, and other lens
voltages are turned off.
The drying gas flow is set to Standby.
If the APCI source is installed, the vaporizer heater is also turned off.
If the installed source is an Agilent Jet Stream source, the sheath gas
heater is also turned off.
The rough pump turns off when the turbo speed is below a certain
threshold.
4 Wait until the instrument completely vents.
5 Close the MassHunter Data Acquisition program and shut down the
computer.
6 Press the power switch located in the lower left front corner of the instrument.
7 Unplug the main power cable behind the service panel.
This action prevents the turbo pump components from overheating.