4. Self-add with password
Login with operator password for measurement; system settings are protected.
Userscanaddthemselvesas‘user’leveloperatorsatlogin,howeveritismanda-
tory for them to select a password. Audit trail applies.
5. Secure
Full security applied: only registered users may login. Users can be registered by
administrators(‘admin’).Audittrailapplies.
11.14.3 Custom security settings (Advanced topic)
To enable full customization for security settings select Custom on the
»Operators»Security screen and press the Customize button, which takes you
to the »Operators»Custom screen
The definitions of the switches are the following:
auto login:
Ifactive,theusermayusethe‘autologin’featuretooperatetheanalyzerwithout
logging into the system. Identification of the user is not forced in this case as the
‘autologin’operatorcanbeusedbyanyone.
Ifactive,thegrantedrightlevelofthe‘autologin’mainlydeterminesthesecurity
level of the system applied.
Tologinwith‘autologin’operator,leaveblanktheloginnameeld(‘EnterOpera-
tor name’) and simply press the Apply button.
self add operators at login:
If active, the user can create a new arbitrary operator name at login (if the login
name does not already exist in the user database).
login without password:
If active, no password usage is automatically forced for each operator. New
operators login into the system without a password (the system does not ask
to set it).
If, however, a password has been already set for a particular operator, that operator
can only be logged in with her/his login name & password combination.
operators on login screen:
If active, a maximum of four operators’ names are displayed on the login
screen.
Supervisor group operators cannot be listed on the login screen.
LIS2 operator list check*):
If active, the operators defined on the LIS can be used also too.
LIS2 only:
If active, only the operators defined on the LIS can be used (except the operators
with supervisor level).
*) LIS2 operators: experimental feature