EasyManuals Logo

Anderson DA2 User Manual

Anderson DA2
56 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #45 background imageLoading...
Page #45 background image
PAGE 45
9. Press ENTER key to display . With the decade box
connected, press ENTER key again to confirm. This active
screen appears showing the measurement reading.
NOTE: During calibration, the analog output is automatically “held” at the last measured
value.
10. Wait for the reading to stabilize which may take up to 30 minutes. Then press ENTER
key. The “PLEASE WAIT” screen may appear if the reading is still too unstable. After
the reading has stabilized, this
static screen appears
showing the “last measured” value.
11. Use arrow keys to adjust the displayed value to exactly match the desired conductivity
value used in the calculation..
12. Press ENTER key to enter the value and complete calibration (“CONFIRM CAL OK?”
screen appears).
13. Press ENTER key again to display the active measurement reading on the “CONFIRM
ACTIVE?” output status screen. When the reading corresponds to the actual typical
process value, press ENTER key again to return the analog output to its active state
(MEASURE screen appears).
This completes ELECTRONIC CAL calibration.
8.4 % Concentration Calibration8.4 % Concentration Calibration
8.4 % Concentration Calibration8.4 % Concentration Calibration
8.4 % Concentration Calibration
After zeroing the sensor (first-time sensor calibration only), calibrate the sensor span using one
of these methods:
CONC CAL Method: This method requires you to immerse the sensor into a prepared
% concentration reference solution of known value, or to keep the sensor installed in
the process while obtaining a process sample. When keeping the sensor installed,
determine the process value by laboratory analysis or comparison reading. In either
case, enter the known reference solution or sample % concentration value.
COND CAL Method: This method requires removing the sensor from the process,
immersing it into a
conductivity reference solution, entering a reference for temperature
compensation temperature, and entering the known linear % per °C slope and conduc-
tivity value of the reference solution. The conductivity reference solution should have
an equivalent, uncompensated value that corresponds with the normal % concentration
value of the process.
1. Depending on the situation, do one of the following:
When Keeping Sensor Installed:
Obtain a sample of the process solution and determine its value using laboratory
analysis or a recently calibrated portable meter.
When Immersing Sensor in Reference Solution:
A. Prepare a % concentration reference solution using your normal method. To
achieve accurate calibration, the reference solution must have the same
chemical composition as the process. Also, its value should be near the typical
measured process value.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Anderson DA2 and is the answer not in the manual?

Anderson DA2 Specifications

General IconGeneral
BrandAnderson
ModelDA2
CategoryTransmitter
LanguageEnglish

Related product manuals