General Information 1-5 Secure Environment Workplace
Site Master User Guide PN: 10580-00252 Rev. K 1-7
1-5 Secure Environment Workplace
This section details the types of memory in the Site Master, how to delete stored user files in
internal memory, and recommended usage in a secure environment workplace.
Site Master Memory Types
The instrument contains non-volatile disk-on-a-chip memory, EEPROM, and volatile DRAM
memory. The instrument does not have a hard disk drive or any other type of volatile or
non-volatile memory.
Disk-On-A-Chip (DOC): DOC is used for storage of instrument firmware, factory calibration
information, user measurements, setups, and .jpg screen images. User information stored on
the DOC is erased by the master reset process described below.
EEPROM: This memory stores the model number, serial number, and calibration data for the
instrument. Also stored here are the user-set operating parameters such as frequency range.
During the master reset process, all operating parameters stored in the EEPROM are set to
standard factory default values.
RAM Memory: This is volatile memory used to store parameters needed for the normal
operation of the instrument along with current measurements. This memory is reset
whenever the instrument is restarted.
External USB Flash Drive (not included with the instrument): This memory may be selected
as the destination for saved measurements and setups for the instrument. The user can also
copy the contents of the internal disk-on-chip memory to the external flash memory for
storage or data transfer. The external Flash USB can be reformatted or sanitized using
software on a PC.
Refer to Chapter 4, “File Management” for additional information on saving and copying files
to the USB flash drive.
Erase All User Files in Internal Memory
Perform a Master Reset:
1. Turn the instrument on.
2. Press the Shift button, then the System (8) button.
3. Press the System Options submenu key.
4. Press the Reset key, then the Master Reset key.
A message box will be displayed to warn the user that all settings will be returned to
factory default values and all user files will be deleted. This is a standard file deletion
and does not involve overwriting existing information.
5. Press the ENTER button to complete the master reset.
6. The instrument will reboot and the reset is complete.
Recommended Usage in a Secure Environment
Set the Site Master to save files to the external USB Flash drive:
1. Attach the external Flash drive and turn the instrument on.
2. Press the Shift button, then the File (7) button.