Instrument Overview 2-4 Interface Screen
Site Master User Guide PN: 10580-00252 Rev. K 2-13
Calibration Symbols
The current calibration status and type is displayed in the upper-left of the screen when in
Cable & Antenna Analyzer mode. See Figure 2-6 on page 2-6. The five status messages are
described next.
Cal Status: ON, Flex
The Site Master has been calibrated with discrete Open, Short, and Load components. This is
a FlexCal calibration indicating it is possible to change the frequency range after calibration.
Cal Status: ON, Standard
The Site Master has been calibrated with discrete Open, Short, and Load components. This is
a Standard calibration indicating it is not possible to change the frequency range after
calibration without performing another calibration.
Cal Status: ON, Flex, Insta
The Site Master has been calibrated with the InstaCal module. This is a FlexCal calibration
indicating it is possible to change the frequency range after calibration.
Cal Status: ON, Standard, Insta
The Site Master has been calibrated with the InstaCal module. The Site Master has been
calibrated with discrete Open, Short, and Load components. This is a Standard calibration
indicating it is not possible to change the frequency range after calibration without
performing another calibration.
Cal Status Off:
The Site Master has not been calibrated.