3-15 Measurements Menu Interference Analyzer (Option 25)
3-64 PN: 10580-00349 Rev. H Spectrum Analyzer MG
RSSI Menu
Key Sequence: Measurements > RSSI
Time Interval: Press the Time Interval submenu key to set the time between
adjacent measurement points. This time may be set from 70 ms to 1 minute.
Time Span: Press the Time Span submenu key to set the overall time span
for the RSSI measurement. This time can be set from zero, to give manual
control of the time span, to a maximum of seven days. After the specified time
span, the measurement is halted. Depending on the time interval selected,
the data will scroll to the left once the trace fills the screen.
Auto Scale: Press the Auto Scale submenu key to automatically set the
reference level and scale factor to place the trace on the screen.
Record On/Off: To store the RSSI data, press the Record On/Off submenu
key to turn on data logging. Each screen full of 551 data points will be stored
as a separate display, and can be saved for up to seven days. The instrument
saves the data in instrument memory and it can be recalled by the Recall
submenu key (File > Recall).
Reset/Restart Measurement: Resets or restarts the measurement.
The RSSI trace is erased and begins anew at the right side of the display.
Back: Returns to the “Measurements Menu” on page 3-41.
Figure 3-58. IA RSSI Menu
Record
On Off
RSSI
Back
Time Interval
70 ms
Time Span
0Ms
Auto Scale
Reset/
Restart
Measurement