LTE Signal Analyzer 5-4 LTE Modulation Measurements
3GPP MG PN: 10580-00234 Rev. T 5-9
Markers
Markers are available for use in the Power vs Resource Block measurement. The markers will
outline the selected RB. Above the two dimensional grid, the specific RB subcarrier/sub-frame
coordinates will be displayed along with its power level.
Constellation
This measurement displays the constellation of the demodulated symbols over the first
sub-frame and a results table of the following measurements.
Ref Signal (RS) Power
Reference Signal power displayed in dBm or Watts. The reference signal is used for downlink
channel estimation.
EVM (rms)
The RMS (%) of all the error vectors, between the reconstructed ideal signals and the
received signals, divided by the RMS value of the ideal signals. When EVM Max Hold is On
(Setup menu) two values are displayed. The first number is the maximum EVM (rms) value
since Reset and the second number is the current measured value. Reset occurs when setup
parameters are changed or by toggling the EVM Max Hold button.
Freq Error
The difference between the measured carrier frequency and the specified carrier frequency is
the frequency error. This number is only as accurate as the frequency reference that is used,
and is typically only useful with a good external frequency reference or GPS.
Carrier Frequency
Carrier Frequency is the measured transmitter operating center frequency.
Sync Signal (SS) Power
Sync signal power displayed in dBm or Watts.
EVM (pk)
The peak (%) of all the error vectors, between the reconstructed ideal signals and the received
signals, divided by the RMS value of the ideal signals. When EVM Max Hold is On
(Setup menu) two values are displayed. The first number is the maximum EVM value since
Reset and the second number is the current measured value.
Freq Error (ppm)
Freq Error displayed in parts per million (ppm).
Cell ID
Cell identifying information sent by the transmitter in the sync signal.
ООО "Техэнком" Контрольно-измерительные приборы и оборудование www.tehencom.com

Table of Contents

Other manuals for Anritsu Spectrum Master MS2720T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Anritsu Spectrum Master MS2720T and is the answer not in the manual?

Anritsu Spectrum Master MS2720T Specifications

General IconGeneral
BrandAnritsu
ModelSpectrum Master MS2720T
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals