Chapter 2 —Fixed WiMAX Signal Analyzer 2-5 Fixed WiMAX Demodulator
WiMAX MG PN: 10580-00236 Rev. D 2-11
EVM versus SubCarrier
This view shows the EVM (rms) values versus OFDM sub-carriers. Eight pilot sub-carriers
are shown in orange, and 192 data sub carriers are shown in yellow. The numerical results
that are displayed in this view are: RCE (rms) in dB, RCE (pk) in dB, EVM (rms) in %,
EVM (pk) in %, Freq Error in Hz, Freq Error in ppm, Carrier Frequency in Hz, Base
Station ID.
Procedure:
1. Setup the measurement as described in “Setup for Fixed WiMAX Demodulator
Measurements” on page 2-8.
2. Press the Measurements main menu key.
3. Press the Demodulator submenu key.
4. Press the EVM vs SubCarrier submenu key to activate the measurement. The sample
image in Figure 2-7 may differ from any image on your instrument.
Figure 2-7. EVM Versus SubCarrier
ООО "Техэнком" Контрольно-измерительные приборы и оборудование www.tehencom.com

Table of Contents

Other manuals for Anritsu Spectrum Master MS2720T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Anritsu Spectrum Master MS2720T and is the answer not in the manual?

Anritsu Spectrum Master MS2720T Specifications

General IconGeneral
BrandAnritsu
ModelSpectrum Master MS2720T
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals