EasyManuals Logo

Applied Biosystems StepOne User Manual

Applied Biosystems StepOne
184 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #132 background imageLoading...
Page #132 background image
DRAFT
September 25, 2007 1:07 am, 4376782_Maintain_the_System.fm
Chapter 6 Maintain the System
Regular Maintenance
Applied Biosystems StepOne
and StepOnePlus
Real-Time PCR Systems
Installation, Networking, and Maintenance Guide
110
Notes
How to Calibrate
a Custom Dye
Materials Required
Centrifuge with reaction plate adaptor
Custom dye(s)
Deionized water
MicroAmp Optical Adhesive Film or MicroAmp
8-Cap Strip (Flat Caps only)
•MicroAmp
Optical Reaction Plate of the size appropriate for your instrument
Pipettors and pipette tips (200-µL and 1000-µL)
Powder-free gloves
Safety glasses
Tubes (2-ml and 10-ml)
Prepare the Custom Spectral Calibration Plate
1. Create a calibration plate for the custom dye:
a. Prepare 1.5 mL of the custom dye at the desired concentration.
b. Pipette 30 µL of the diluted custom dye to all wells of a reaction plate.
c. Seal the wells of the reaction plate using an adhesive film or optical flat caps.
d. Vortex the calibration plate for 5 sec.
e. Centrifuge calibration plate for 2 min at less than1500 rpm.
IMPORTANT! The calibration plate must be well mixed and centrifuged.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems StepOne and is the answer not in the manual?

Applied Biosystems StepOne Specifications

General IconGeneral
BrandApplied Biosystems
ModelStepOne
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals