SECTION 7 DIAGNOSTICS AND TROUBLESHOOTING
______________________________________________________________________
7-8
E45 Message:
Operator Action:
No Updates From BIS Engine
No processed data from BIS Engine in the last 10 seconds. If
error persists, monitor may require service.
E46 Message:
Operator Action:
BIS Engine Not Responding
BIS Engine is not performing commands within the allowed
time limits. If error persists, monitor may require service.
E47, E74, E81
Message:
Operator Action:
Data Memory Write Error
Problem writing data to non-volatile memory. Monitor
requires service, but should continue to operate properly
(data memory storage time will be reduced).
E48 – E65 (See E35)
Software Error
E66 Message:
Operator Action:
Printer out of Paper
Check the printer. Make sure that it has paper. If the error
persists, the printer requires service.
E67, E82, E85
Message:
Operator Action:
Printer Error
This indicates a printer hardware error. If the error persists,
the printer requires service.
E68 (not used)
E69 – E73 (See E35)
Software Error
E74 (See E47)
Data Memory Write Error
E75 (See E35)
Software Error
E76 Message:
Operator Action:
Illegal Sensor for Demo
A demo device is attached to the serial port, but the sensor
connected through the DSC is not a valid demo sensor.
E77 Message:
Operator Action:
Simulator Connected
No action required. This message indicates that a demo
device is attached to the serial port of the A-2000 and the
A-2000 is not necessarily processing patient information
E78 – E80 (See E35)
Software Error
E81 (See E47)
Data Memory Write Error
E82 (See E67)
Printer Error