SECTION 7 DIAGNOSTICS AND TROUBLESHOOTING
______________________________________________________________________
7-9
E83 Message:
Operator Action:
Printer Head is Up
The print head on the printer has been lifted. Return it to the
closed position. If the error persists, the printer requires
service.
E84 Message:
Operator Action:
Printer Disconnected
The printer is no longer connected to the A-2000. Check the
connection. If the error persists, the printer requires service.
E85 (See E67)
Printer Error
E86 Message:
Operator Action:
Printing Disabled
The A-2000 is running on battery backup. When power is
restored to the A-2000, printing will again be available.
E87- E89 (See E35)
Software Error
E90 Message:
Operator Action:
Printer Connected
The printer has been successfully connected to the A-2000.
E91 Message:
Operator Action:
Event Marking Disabled
The ability to mark an event has been temporarily disabled. It
will be re-enabled when event data extraction has been
completed (less than 10 minutes).
E92 (not used)
E93 Message:
Operator Action:
Sensor Ground Fault
Disconnect and examine sensor; replace sensor if necessary.
If problem continues, connect new DSC, restart monitor.
E94 Message:
Operator Action:
Sensor Overcurrent
Disconnect and examine sensor; replace sensor if necessary.
E95-E108 Message:
Operator Action:
Invalid Sensor
Replace sensor if necessary. The sensor is unusable for one
of the following reasons:
E95 – The sensor has been connected too many times
E96-E102, E107 – The sensor contains invalid information
E103-E106, E108 – The sensor information could not be
read.
E109 Message:
Operator Action:
Sensor used > 24 hours
The sensor has been in place for more than 24 hours.
Disconnect and replace sensor.
E110 (See E35)
Software Error