19
GENERAL
Resolution:
0.1W / 1W
Accuracy :
± (5 % of reading + 15
counts)
V > 60V & PF > 0.1
Power Factor High/Low Limit
Setting
Range:
0.000 - 1.000
Resolution:
0.001
Power Factor Display
Range:
0.000 - 1.000
Resolution:
0.001
Accuracy :
W / VA, Calculated and
displayed to three
significant digits
Delay Timer Setting
Range:
0.5-999.9 seconds
Resolution:
0.1
Dwell Timer Setting
Range:
0, 0.1-999.9
(0=continuous)
Resolution:
0.1
Timer display
Range:
0.0 999.9 seconds
Resolution:
0.1 second
Accuracy:
± (0.1% of reading + 0.05
seconds)
Why use the term “Counts”?
Associated Research publishes some specifications using COUNTS which allows us to provide
a better indication of the tester’s capabilities across measurement ranges. A COUNT refers to
the lowest resolution of the display for a given measurement range. For example, if the
resolution for voltage is 1V then 2 counts = 2V.