Refer to section 7.5, Non Volatile Memory for more information.
SS <step number>
Selects the active selected step to load into RAM. The step must first be selected
before any specific parameters can be edited.
ADD <test p1,p2,p3…>
This command edits all parameters in a step. Parameters will be edited at the step
location that has been selected. See the command summary tables below to see the
specific test type for each of these commands
The parameter <test> indicates the test type. The values LLT and RUN must be
used. The parameters <p1,p2> etc. indicate the individual settings for each parameter
of the test. All parameters must be included with the command and should appear in
the same order that is shown in the table below. Also, like the individual parameter
editing commands, the unit should not be included with the value; only the numeric
value should be included in the command string. When the scanners are being used
they should be appended to the end of the string.
The list of parameters can also be found in the default parameters section of the
manual, or refer to Section 7.4.2. Test Parameter Editing Commands and
Companion Queries for the proper values.
The parameter values for file editing commands should use complete text (i.e. "ON"
and "OFF") and not use the coded values that are associated with the test parameter
setting commands discussed in Section 7.4.2. Test Parameter Editing Commands
and Companion Queries. The LS? companion command will also list all parameters
in complete text in the order as they appear in the following table, preceded by the
step number.