ATI Series D12 Toxic Gas Transmitter with H10 Smart Sensor
Revision N (7/15) 30
Startup Review Sequence
When the transmitter starts, the display cycles through a series of pages to review the configuration of
the transmitter, sensor, and generator. Alarms are inhibited, and the output of the transmitter is held at
4.0 mA (17.4mA for Oxygen sensors). This state is maintained for 5 minutes to provide time for the
sensor readings to stabilize.
Figure 25. Startup review sequence
AUTOTEST
Status=OFF
►Hold Next Skip
ATI Model D12
Tox Gas Monitor
►Hold Next Skip
Version Info
HW=X.XX SW=X.XX
MM/DD/YY
►Hold Next Skip
Transmitter
Self Check
Pass
►Hold Next Skip
System Clock
Monday
MM/DD/YY HH:MM
►Hold Next Skip
24.2 H2S
Range= 10 PPM
To 200 PPM
►Hold Next Skip
PPM
H2S
►Hold Next Skip
PPM
H2S
►Verify Range
24.2 H2S
Range= 10 PPM
To 200 PPM
►Verify Sensor
PPM
H2S
►Hold Next Skip
PPM
H2S
►Hold Next Skip
PPM
H2S
►Verify A-Alarm
Datalog Status
Logging is OFF
0 days of 11
►Hold Next Skip
24.2 H2S
10 PPM 1.00mA
0.00Hrs of 3.00
►Hold Next Skip
Gas Generator
Not Installed
►Hold Next Skip
AUTOTEST
Status=READY
09/12/06 10:15
►Hold Next Skip
Gas Generator
Not Installed
Cannot Autotest
►Cancel Fault
Gas Generator
Incompatible
Cannot Autotest
►Cancel Fault
If the auto-test state is READY and the generator is
not installed, or is not compatible with the sensor type
and range, select “Cancel Fault” to force the auto-test
state to OFF. This is the default action if no action is
taken within 5 minutes.
PPM
H2S
►Verify C-Alarm
PPM
H2S
►Hold Next Skip
PPM
H2S
►Verify W-Alarm
Information
Self-
setting
Sensor
information
Programmed full
scale range
Caution alarm set
point and direction
Warning alarm set
point and direction
nd
direction
Auto-test status
Sensor Status
Not Installed
►Hold Next Skip
Startup Review
PPM
H2S
Main Display
Displays the date and time of
the next scheduled Auto-test
(if Status is READY). Date is
automatically adjusted when
past due, or within 24 hours
(see Auto-test)
Datalog Status
Logging is ON
11 days of 11
Datalog Status
Logging is ON
11 days of 11
►Clear Datalog
Datalog is cleared
after installing new
type of sensor.
OR
OR
OR
OR
OR
OR
OR