EasyManua.ls Logo

ATI Technologies d12 series - Calibration of Duct-Mount H10 Sensors; Figure 38. Calibration of Duct-Mount H10 Sensors (ATI-0664)

ATI Technologies d12 series
92 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ATI Series D12 Toxic Gas Transmitter with H10 Smart Sensor
Revision N (7/15) 41
Calibration of Duct-Mount H10 Sensors
To Zero and Span calibrate a duct-mounted H10 sensor; pull the sensor out of the duct until the
calibration locking pin clicks into the retaining slot. By pulling back slowly, the retaining pin will drop
into place and automatically locate the sensor just behind the gas inlet port. Connect a length of tubing
from the gas regulator to the barb fitting, and then follow the zero and span calibration procedures
detailed earlier.
Figure 38. Calibration of duct-mount H10 sensors (ATI-0664)
For best results, pressure in the pipe
or duct should be neutral, or slightly
negative. Positive pressure may restrict the flow of calibration gas to the
sensor, resulting in an inaccurate calibration.

Table of Contents

Related product manuals