Preparation before Use
20
[3] Preparation before Use
3-1.
Assembly
Be sure to secure the incubator and a stand using the incubator retainer. Failure to do so may
cause the incubator to tip over if a force is applied while the hood is open, or to fall due to
shock received during transportation. Never operate the incubator until it has been firmly se-
cured.
The incubator is heavy and should be assembled and moved by at least two people.
The incubator is packed when shipped. When using the incubator placed on the HL or cabinet stand, follow
the procedure below to assemble the stand.
3-1-1.
For HL Stand (Option)
(1) Mount the incubator on the high-low stand.
(2) Be sure that the rubber feet of the incubator are
placed on the incubator retainer before tightening
the lever of the retainer to secure the incubator.
Lever
Incubator retainer
3-1-2.
For Cabinet Stand (Option)
(1) Mount the incubator on the cabinet stand.