1
INTRODUCTION
Thank you for purchasing IncuArch.
This Operation Manual deals with the specifications, operation and maintenance of the IncuArch. Atom is by no
means responsible for any malfunction arising from a user ignoring the instructions for operation and mainte-
nance described in this Manual as well as any accident attributable to repair by someone other than technical
personnel belonging to or authorized by Atom.
Read this Manual carefully and familiarize yourself with its contents before operating the unit. Keep this Manual
where it is readily accessible for reference when needed. If any technical problems should arise, please contact
your local Atom representative.
The unit is shipped without being disinfected. Be sure to clean and disinfect the unit before using
it for the first time after purchase.
The product and its parts that are past their service lives should be disinfected before being dis-
posed of in accordance with the applicable waste management laws.
This product maintains the ambient temperature around an infant by drawing in the air outside
and warming it. It does not have the capability to cool the air drawn from outside. Be sure to set
the incubator air temperature at least 3°C higher than the ambient temperature. In particular, if a
phototherapy unit, a heated humidifier, or some other heat-generating device is used with the in-
cubator, set the incubator air temperature at least 5°C higher than the ambient temperature. If a
lower setting is selected, the incubator may not be controlled correctly.
This product is categorized as a portable medical device. It is intended for installation in an ambu-
lance or rotorcraft. Make sure to perform test operation when installing it to confirm that there is
no problem with the operation of the product or the vehicle where it is installed. For the device
characteristics, see "[15] Device Characteristics - EMC Level and Classification."
INTENDED USE
The IncuArch is a closed-care transport incubator for neonatal and premature infants and is intended to transport
low birth weight infants and seriously ill neonates.