MAINTENANCE
Cleaning and Disinfection
71
[10] Cleaning and Disinfection
The heaters remain very hot after use. Allow the heaters to cool down sufficiently before clean-
ing and disinfecting them in order to avoid burns.
A fire and explosion hazard exists when performing cleaning in an oxygen-enriched environ-
ment. Turn off the oxygen supply and disconnect the oxygen supply hose from the incubator
before cleaning the incubator.
The unit is shipped without being disinfected. Be sure to clean and disinfect the unit before us-
ing it for the first time after purchase.
To prevent infections, disinfect the unit at regular intervals. The intervals to disinfect and disinfec-
tion methods vary depending on the usage frequency and the environment in which the unit is
used. Medical institutions that use the unit should determine appropriate disinfection intervals
and methods.
See the accompanying document of the disinfectant you are going to use for appropriate con-
centration, contact time and handling. Follow the directions for use and dosage and precau-
tions given in the document.
Be sure to dilute disinfectants, such as Osvan, Hyamine, and Hibitane. NEVER use undiluted
solution.
Do not use any abrasive cloth, cleaner, alcohol, acetone or other types of solvent for cleaning
or disinfecting.
Turn the power switch off and unplug the power plug before cleaning and disinfecting the unit.
Clean and disinfect the unit whenever it is used for a different infant.
Clean and disinfect the unit whenever you notice any dirt or contamination which may intro-
duce an infection.
After cleaning and disinfecting, reassemble the removed parts correctly and check that the unit
operates normally.
After cleaning and disinfecting, remove the disinfectant solution thoroughly from the incubator
before using it.
Provide a soft clean cloth and a disinfectant solution suitable for cleaning and disinfection.