OPERATION
Oxygen Supply
53
(6) When the calibration is successfully completed,
“Calibration Completed” will be displayed.
To return to the menu screen, press the
switch once. To return to the main screen, press
the switch again.
z
If the display shows “Calibration Error, the oxy-
gen sensor may be faulty. Replace it with a new
one.
z
If the display shows “Calibration Error”, the oxy-
gen concentration display will be "---" until the cal-
ibration is successfully completed.
(7) Insert the oxygen sensor module into the oxygen
sensor inlet.
z
Be sure to calibrate the oxygen sensor before use.
z
To calibrate the oxygen sensor, it must be placed
in ambient air (21% oxygen).
z
Be sure to calibrate the oxygen sensor with the
sensor in the oxygen sensor module facing down-
ward.
Oxygen sensor module
Oxygen
sensor inlet
7-2-2.
Setting the Upper and Lower Alarm Limits for Oxygen Level
The unit allows you to set the upper and lower alarm limits for oxygen level. Audible and visual alarms will
be activated if any of the alarm limits are violated (see “[12] Alarms”).
(1) When you press the
switch, the cursor control
screen appears.