MAINTENANCE
Maintenance and Inspection
79
11-1- 2.
Inspection Checklist - During Use
Device Name IncuArch
Date of
Inspection
Serial No. Inspector
Check the following points periodically while using the unit.
No. Inspection Item to check Yes/No
1 Main body
(1) Are the main body and hood free of damage and not deformed in
any way?
(2) Is the power cord securely connected?
(3) Are the main body and stand securely locked?
(4) Are the admittance panels and snap-open access ports securely
locked?
(5) Are the air inlets and outlets free of obstacles that may block air
circulation?
* Check that no objects are placed in any spaces other than the
mattress in the incubator.
2 Operation
(1) Is the incubator air temperature controlled as specified?
* Be sure to set the incubator air temperature at least +3°C higher
than the ambient temperature. If a phototherapy unit or a heated
humidifier is used in combination with the incubator, set the
incubator air temperature at least +5°C higher than the ambient
temperature.
(2) Battery
inspection:
When using an external DC power supply
Does the power supply switch to battery mode
when shut off the DC power supply?
When running on battery power, is there enough
charge remaining?