Maintenance and Inspection
78
11-1- 1.
Inspection Checklist - Before Use
Device Name IncuArch
Date of
Inspection
Serial No. Inspector
Check the following points every time you are going to use the unit.
No. Inspection Item to check Yes/No
1 Main body
(1) Are there any appearance-related abnormalities?
(2) Are the main body and stand securely locked?
(3) When the admittance panels and snap-open access ports are fully
closed, do they stay locked in place when they are pushed from
the inside?
(4) Is each part of the unit correctly assembled?
2
Operation
Checking for
operation when
the power is
turned on
(1) Is the power cord securely connected and has the AC power
indicator lit up?
* When there is remaining battery charge, the AC power indicator
is activated even if the power cord is not connected. Check that
the power cord is securely connected.
(2) When the power switch is turned on, will the startup sound, and
the main screen will be displayed after the startup screen is ap-
peared?
(3) Is the measured value displayed when you connect and attach the
skin temperature probe?
(4) Is the measured value displayed when you connect and attach the
SpO
2
sensor?
(5) Is the measured value displayed after you connect the oxygen
sensor module and perform calibration?
(6) Does the lighting lamp light up?
(7) Battery
inspection:
Does the power supply switch to battery mode
when the power plug is disconnected from the
power outlet?
Does the battery status indicator show a white
frame with three lit cells after the battery has
been charged?
When using an external DC power supply Does
the power supply switch to battery mode when
shut off the DC power supply?
When running on battery power, is there enough
charge remaining?