MAINTENANCE
Cleaning and Disinfection
75
(5) Hold both ends of the middle board, raise it and
then remove.
z
The heater and its surrounding surfaces may be
hot enough to cause burns. Avoid touching them
until they have cooled down.
Middle board
(6) Remove the fan by pulling it up.
z
When placing the cleaned fan back into place, be
sure to push the fan onto the shaft as far as it will
go. An incorrectly installed fan may cause abnor-
mal noise or malfunction.
Fan
(7) When all parts have been removed by following
the procedures above, the conditioning chamber
comes into view.
Clean the inside of the conditioning chamber using
a soft cloth that has been soaked in a disinfectant
solution and wrung out.
(8) Disinfect the removed parts.
Parts that are to
be cleaned by
immersing them
in a disinfectant
solution
Fan
Parts that are to
be cleaned
using a soft cloth
that has been
soaked in a
disinfectant
solution and
wrung out
Mattress
Guard
Mattress platform
Packing for main body
Middle board
(9) After cleaning and disinfecting, allow each part to
dry sufficiently and reassemble all parts removed
in reverse order.