Alarms
86
[12] Alarms
The unit provides the following alarm features. If an alarm condition should occur, check for a
possible cause of the alarm and take the correct measures. If the alarm is caused most likely
by a failure, the unit needs to be repaired. Indicate on the incubator that it is out of order and
contact your local Atom representative.
After the power has been turned on or a setting value has been changed, the set temperature
alarm does not function for 50 minutes. If you changed the set temperature during transport,
watch the infant condition carefully. However, when the incubator temperature becomes ±3°C
of the set value within 50 minutes, the set temperature alarm activates at that time.
Cat-
egory
Alarm
name
Condition causing
the alarm
Icon
displayed
When an
alarm is
activated:
Cancellation of alarm
Priority
level
✽1
Alarm
is silenced
Temperature & Heat
Excessive
tempera-
ture alarm
This alarm activates
before the incubator air
temperature exceeds
39°C.
The incubator
heater is
deactivated.
The heater is automatically
reactivated when the
temperature is reduced to
the normal level, however,
the audible alarm stays on.
The alarm is canceled by
pressing the silence switch.
Middle
10 min.
Set
tempera-
ture alarm
This alarm activates if
the incubator air
temperature differs
from the set tempera-
ture by ±3°C or more.
✽2
Even if the
temperature
exceeds -3°C
or more,
control is
maintained.
If the
temperature
exceeds +3°C
or more, the
heater is
turned off.
The alarm is automatically
canceled when the tempera-
ture difference is less than
3°C.
Low 15 min.
Skin
tempera-
ture probe
alarm
This alarm activates if
the skin temperature
probe is electrically
open or short-circuited.
Skin tempera-
ture display
“- - -”
The alarm is automatically
canceled when the probe is
replaced with a good one.
Low 15 min.