88 300-002547-00 rev7
Parts Replacement
The Monarch Platform does not have any user-serviceable parts with the
exception of the following system accessories:
Monarch Controller.
Navigation Patient Sensors.
Bronchoscope Patient Introducer Mount.
Field Generator Mount.
All replacement parts must be provided by Auris Health, Inc. Contact Auris
Customer Care for assistance. Use of third-party components or modifications of
the Monarch Platform may result in unexpected behavior, creates safety risks,
and voids product warranties. Auris Health, Inc. is not responsible for any
damages resulting from use of unauthorized parts or modifications.
Replacement of Monarch Controller
1. Open the Monarch Tower drawer.
2. Disconnect the Monarch Controller connector from the drawer connector.
3. Connect replacement Monarch Controller to the drawer connector.
4. Press the pause/play button on the Controller to ensure the controller is
active.
Replacement of the Navigation Patient Sensor Cable
1. Ensure that the Navigation Patient Sensor is detached from the patient and
the Monarch Platform is not in a procedure.
2. Disconnect the Navigation Patient Sensor cable from the Tower connector.
3. Plug in the replacement Navigation Patient Sensors to the Tower.
4. Follow the instructions in System Setup (chapter 3) to attach the Navigation
Patient Sensors to the patient and confirm that the sensors are detected in
the Navigation setup menu.
Replacement of the Bronchoscope Patient Introducer Mount
1. Ensure that the Monarch Platform is not in a procedure.
2. Remove the Bronchoscope Patient Introducer from the Patient Introducer
Mount, if needed.
3. Attach Bronchoscope Patient Introducer Mount to the bed rail.
4. Remove the Bronchoscope Patient Introducer Mount from the bed rail.
5. Follow the instructions in Chapter 3: System Setup to install the replacement
Bronchoscope Patient Introducer Mount on the bed rail.

Table of Contents