300-002547-00 rev7 89
Replacement of the Field Generator Mount
1. Ensure that the Monarch Platform is not in a procedure.
2. Remove the Monarch Navigation Field Generator from the Monarch
Navigation Field Generator Mount, if installed on the bed rail.
3. Loosen the clamp securing the Monarch Navigation Field Generator Mount to
the bed rail.
4. Remove the Monarch Navigation Field Generator Mount from the bed rail.
5. Follow the instructions in Chapter 3: System Setup to install the replacement
Monarch Navigation Field Generator Mount to the bed rail.
Recommended Spare Components
Auris recommends that the following spare components and accessories are
readily available as necessary:
• Controller.
• Patient Sensors.
• Disposable accessories.