300-002547-00 rev7 105
ID Description Message
5 Nav Error A system error has occurred and all robotic
functions have been disabled. Press OK to
continue. If problem persists, contact Auris
support.
6 Nav Error A system error has occurred and all robotic
functions have been disabled. Press OK to
continue. If problem persists, contact Auris
support.
7 Nav Error A system error has occurred and all robotic
functions have been disabled. Press OK to
continue. If problem persists, contact Auris
support.
8 System Error A system error has occurred and all robotic
functions have been disabled. Press OK to
continue. If problem persists, contact Auris
support.
9 System Error A system error has occurred and all robotic
functions have been disabled. Press OK to
continue. If problem persists, contact Auris
support.
100 System Error A system error has occurred and all robotic
functions have been disabled. Press OK to
continue. If problem persists, contact Auris
support.
101 Auto Bring-up failure A system error has occurred. All robotic functions
have been disabled. To resolve, unload then
reload the Bronchoscope System and click OK to
acknowledge the fault. Then, re-initialize
navigation at the main carina.
103 Configuration Server Failure A system error has occurred. All robotic functions
have been disabled. Unload the Bronchoscope
System and reboot the system. If problem
persists, contact Auris Technical support.

Table of Contents