106 300-002547-00 rev7
ID Description Message
104 System Error A system error has occurred and all robotic
functions have been disabled. Press OK to
continue. If problem persists, contact Auris
support.
105 The Cart is disconnected from
the Tower. Please make sure
the Cart is connected to the
Tower.
A system error has occurred. All robotic functions
have been disabled. Unload the Bronchoscope
System and reboot the system. If problem
persists, contact Auris support.
107 SlaveApp Restart detected A system error has occurred. All robotic functions
have been disabled. To resolve, unload then
reload the Bronchoscope System and click OK to
acknowledge the fault. Then, re-initialize
navigation at the main carina.
108 EndoConsole restart detected A system error has occurred. All robotic functions
have been disabled. To resolve, unload then
reload the Bronchoscope System and click OK to
acknowledge the fault. Then, re-initialize
navigation at the main carina.
200 Sheath insertion force limit Sheath insertion force limit has been reached.
Check the sheath and arm alignment to the
patient introducer, check the device lubrication,
and clear the insertion pathway. Press OK to
continue.
201 Scope insertion force limit Bronchoscope insertion force limit has been
reached. Retract and visually inspect the device
for damage and kinks. Lubricate as necessary
prior to reinsertion. Press OK and initialize
navigation at the main carina.
202 High sheath insertion force has been detected.
203 High bronchoscope insertion force has been
detected.
204 Bronchoscope insertion limit has been reached.