300-002547-00 rev7 107
ID Description Message
205 Bronchoscope retraction limit has been reached.
206 Sheath insertion limit has been reached its limit.
207 Sheath retraction limit has been reached.
208 Sheath insertion limit has been reached.
209 Bronchoscope System loading
error
The Bronchoscope System failed to load
properly. Unload and reload the device.
210 Tensioning error The Bronchoscope System failed to load
properly. Unload and reload the device.
212 Bronchoscope articulation limit has been
reached.
213 Robot collision An arm collision has been detected, remove
obstructions to proceed. Press OK to continue.
214 Potential Collision Detection A system error has occurred. All robotic functions
have been disabled. To resolve, unload then
reload the Bronchoscope System and click OK to
acknowledge the fault. Then, re-initialize
navigation at the main carina.
215 Wire break detection Bronchoscope System wire breakage has been
detected. Unload and replace the device.
217 Instrument Still Loaded A system error has occurred. All robotic functions
have been disabled. To resolve, unload then
reload the Bronchoscope System and click OK to
acknowledge the fault. Then, re-initialize
navigation at the main carina.